RapidFire Clip, Item Number 95-0154 - Product Usage: Fixation of Craniofacial bone.

Class I - Dangerous
🏥 Medical Devices Recalled: November 25, 2020 Biomet Other Medical Devices Nationwide

What Should You Do?

  1. Check if you have this product:
    Lots 080580 231760 359160 359170 381910 382870 382900 382970 383010
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Biomet, Inc.
Reason for Recall:
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

RapidFire Clip, Item Number 95-0154 - Product Usage: Fixation of Craniofacial bone.

Product Codes/Lot Numbers:

Lots 080580 231760 359160 359170 381910 382870 382900 382970 383010

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0626-2021

Related Recalls