NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Nobel Biocare Usa Llc
Reason for Recall:
Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Product Codes/Lot Numbers:

Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0658-2022

Related Recalls

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

Jul 19, 2019 Implants & Prosthetics Nationwide View Details →

Packaged dental implant healing abutments were found to contain the wrong size device.

May 28, 2019 Implants & Prosthetics Nationwide View Details →

Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.