Philips Respironics V60 Ventilator Part Number 1053617
Class I - DangerousWhat Should You Do?
- Check if you have this product: The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1053617 Serial Numbers/UDI Code: MX0016593 / (01)00884838020054(21)MX0016593 MX0016797 / (01)00884838020054(21)MX0016797 MX0016824 / (01)00884838020054(21)MX0016824 MX0016827 / (01)00884838020054(21)MX0016827 MX0016830 / (01)00884838020054(21)MX0016830 MX0016839 / (01)00884838020054(21)MX0016839 MX0016840 / (01)00884838020054(21)MX0016840 MX0016841 / (01)00884838020054(21)MX0016841 MX0016845 (01)00884838020054(21)MX0016845 MX0016853 (01)00884838020054(21)MX0016853 MX0016867 (01)00884838020054(21)MX0016867 MX0016870 (01)00884838020054(21)MX0016870 MX0016871 (01)00884838020054(21)MX0016871 MX0016872 (01)00884838020054(21)MX0016872 MX0016875 (01)00884838020054(21)MX0016875 MX0016878 (01)00884838020054(21)MX0016878
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Respironics California, LLC
- Reason for Recall:
- A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips Respironics V60 Ventilator Part Number 1053617
Product Codes/Lot Numbers:
The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1053617 Serial Numbers/UDI Code: MX0016593 / (01)00884838020054(21)MX0016593 MX0016797 / (01)00884838020054(21)MX0016797 MX0016824 / (01)00884838020054(21)MX0016824 MX0016827 / (01)00884838020054(21)MX0016827 MX0016830 / (01)00884838020054(21)MX0016830 MX0016839 / (01)00884838020054(21)MX0016839 MX0016840 / (01)00884838020054(21)MX0016840 MX0016841 / (01)00884838020054(21)MX0016841 MX0016845 (01)00884838020054(21)MX0016845 MX0016853 (01)00884838020054(21)MX0016853 MX0016867 (01)00884838020054(21)MX0016867 MX0016870 (01)00884838020054(21)MX0016870 MX0016871 (01)00884838020054(21)MX0016871 MX0016872 (01)00884838020054(21)MX0016872 MX0016875 (01)00884838020054(21)MX0016875 MX0016878 (01)00884838020054(21)MX0016878
Distribution:
Distributed in: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0662-2022
Related Recalls
Philips Respironics V680 Ventilator (All Models, including: P/N 1103044 (850011)), *Distributed outside the United States
Respironics California
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Philips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137276)
Respironics California
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Philips Respironics V60 Ventilator (All Models, including: 1053613, 1053614, 1053615, 1053616, 1053617, 1053618, 1076709, 1076716, 1076717, 1137292, DU1053617, U1053617, R1053618, R1076709)
Respironics California
Ventilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.