Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma Catalog number: 07026773190

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Anti-HAV IgM APP-file (version 06.05-101)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Roche Diagnostics Operations, Inc.
Reason for Recall:
Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10).
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma Catalog number: 07026773190

Product Codes/Lot Numbers:

Anti-HAV IgM APP-file (version 06.05-101)

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0669-2020