Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head
Class I - DangerousWhat Should You Do?
- Check if you have this product: REF 36693646 Serial Numbers: 09230967, 09229086, 10230508 REF 36693848 Serial Numbers: 08228745, 10230509, 11235075 REF 36694050 Serial Numbers: 09229091, 09230501 REF 36694252 Serial Numbers: 09229131, 10230510 REF 36694454 Serial Numbers: LLALMB161A, 09230503 REF 36694656 Serial Numbers: 09230505, 11230847 REF 36694858 Serial Numbers: 09229167, 10230511 REF 36695060 Serial Numbers: 09229169, 09230506 REF 36695262 Serial Numbers: 09230507, 09229170 REF 36695464 Serial Numbers: 08228747, 10230512
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- MicroPort Orthopedics Inc.
- Reason for Recall:
- Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected tissue reaction to metal debris over one percent for modular head CONSERVE(R) systems and DYNASTY(R) metal acetabular liner systems.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Lineage Shell: REF 36693646, REF 36693848, REF 36694050, REF 36694252, REF 36694454, REF 36694656, REF 36694858, REF 36695060, REF 36695262, REF 36695464 For use with the conserve metal head
Product Codes/Lot Numbers:
REF 36693646 Serial Numbers: 09230967, 09229086, 10230508 REF 36693848 Serial Numbers: 08228745, 10230509, 11235075 REF 36694050 Serial Numbers: 09229091, 09230501 REF 36694252 Serial Numbers: 09229131, 10230510 REF 36694454 Serial Numbers: LLALMB161A, 09230503 REF 36694656 Serial Numbers: 09230505, 11230847 REF 36694858 Serial Numbers: 09229167, 10230511 REF 36695060 Serial Numbers: 09229169, 09230506 REF 36695262 Serial Numbers: 09230507, 09229170 REF 36695464 Serial Numbers: 08228747, 10230512
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0686-2017
Related Recalls
EVOLUTION MP FEMORAL COMPONENT CS/CR POROUS SIZE 5 RIGHT PRIMARY, REF: EFSRP5PR; EVOLUTION MP FEMORAL COMPONENT CS/CR NON-POROUS SIZE 6 LEFT PRIMARY, REF: EFSRN6PL Knee prosthesis component
MicroPort Orthopedics
Femoral implant porous size 5 right primary and femoral implant non-porous size 6 left primary have correct markings on devices, but may have incorrect packaging.
EVOLUTIONΒΏ MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cemented use only.
MicroPort Orthopedics
One confirmed incident has been received that EFSRN4PR, lot number MP1982170, Size 4 Evolution MP Non-Porous Femoral Right was opened during surgery and contained part number EFSRN3PL, lot number MP1982097, Size 3 Evolution MP Non-Porous Femoral Left in the packaging.
EVOLUTION MP Tibial Bases, Size 7 LEFT, REF ETPKN7SL, knee tibial base
MicroPort Orthopedics
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.