ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All serial numbers of ADVIA 560 Hematology Systems
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Healthcare Diagnostics, Inc.
Reason for Recall:
Siemens is investigating an issue which may cause an incorrect result to be reported. Siemens received two reports of multiple discordant records for the same Sample ID in the ADVIA 560 Hematology System database that occurred during the installation of the systems. The database should only contain one record of a Sample ID number for any given time and date. If there are multiple records for the same Sample ID, it is possible that multiple results may be manually or automatically sent to the Laboratory Information System (LIS), printed or displayed on the results report screen.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.

Product Codes/Lot Numbers:

All serial numbers of ADVIA 560 Hematology Systems

Distribution:

Distributed in: US, CA, CO, FL, IL, MI, NJ, NY, OH, TX, WA, WI, WY

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0723-2017

Related Recalls

Class I - Dangerous

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Mar 18, 2026 Other Medical Devices Nationwide View Details →

A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.

Feb 23, 2026 Other Medical Devices Nationwide View Details →