Magnesium on RX Imola analyser IVD

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    reagent: MG3880 analyser: RX4900
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Randox Laboratories, Limited
Reason for Recall:
According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Magnesium on RX Imola analyser IVD

Product Codes/Lot Numbers:

reagent: MG3880 analyser: RX4900

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0725-2017