iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI 04015630971749, Lot Numbers: Y11834, Y24245
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ventana Medical Systems Inc
- Reason for Recall:
- Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
iView DAB IHC Detection Kit, Ventana part Number 760-091, Roche GMMI 05266157001 in vitro diagnostic
Product Codes/Lot Numbers:
UDI 04015630971749, Lot Numbers: Y11834, Y24245
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0731-2018
Related Recalls
Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Ventana Medical Systems
Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001
Ventana Medical Systems
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.
Ventana HE 600 System, automated slide preparer, for use in laboratories.
Ventana Medical Systems
There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire