Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot #/Expiration Date: 19A0630/January 03, 2029 19B0329 /January 21th, 2029 19C0734 /March 03, 2029
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medicrea International
Reason for Recall:
It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only

Product Codes/Lot Numbers:

Lot #/Expiration Date: 19A0630/January 03, 2029 19B0329 /January 21th, 2029 19C0734 /March 03, 2029

Distribution:

Distributed in: US, NY, CO, OH, AZ, FL, TX, PA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0753-2020

Related Recalls

Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.