Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: (01)07290108240061(17)211031(10)A839(90)3609812 5(91)6070111 Lot Numbers: 7153, 7154, 7634, 7688, 7752, 7909, 7955, 8152, 8153, 8154, 8232, 8233, 8234, 8235, 9028, 9175, 9029, 9176, 9177, 9178, 9367, 9368, 9369, 9370, 9654, 9656, 9655, 9918, 9710, 9711, 9712, 9713, 9839, 9840, 9983, 9984, 9992, 9993, 9994, 9995, A123, A124, A790, A805, A806, A791, A792, A793, A794, A795, A796, A797, A807, A808, A809, A810, A834, A835, A836, A837, A838 & A839
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- WEST PHARMA. SERVICES IL, LTD
- Reason for Recall:
- Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women. There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only
Product Codes/Lot Numbers:
UDI: (01)07290108240061(17)211031(10)A839(90)3609812 5(91)6070111 Lot Numbers: 7153, 7154, 7634, 7688, 7752, 7909, 7955, 8152, 8153, 8154, 8232, 8233, 8234, 8235, 9028, 9175, 9029, 9176, 9177, 9178, 9367, 9368, 9369, 9370, 9654, 9656, 9655, 9918, 9710, 9711, 9712, 9713, 9839, 9840, 9983, 9984, 9992, 9993, 9994, 9995, A123, A124, A790, A805, A806, A791, A792, A793, A794, A795, A796, A797, A807, A808, A809, A810, A834, A835, A836, A837, A838 & A839
Distribution:
Distributed in: US, MO
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0767-2019
Related Recalls
Vial2BAG DC 20mm: 1) REF 6070112 & 2) REF 6070112C, Single use only, Rx Only
WEST PHARMA. SERVICES IL
Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag DC 20mm is recalled due to similarities with 13mm Vial2Bag DC.
Vial2BAG Transfer Device, REF 6070104, Single use only, Rx Only
WEST PHARMA. SERVICES IL
Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag is recalled due to similarities with 13mm Vial2Bag DC.