Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 00763000178338, Serial Number RSL604096S
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Inc.
Reason for Recall:
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1

Product Codes/Lot Numbers:

UDI/DI 00763000178338, Serial Number RSL604096S

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0779-2024

Related Recalls

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Jun 18, 2025 Infusion Pumps Nationwide View Details →

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Jun 18, 2025 Infusion Pumps Nationwide View Details →