AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the disc space with a fixed engagement between the two components to permit a controlled positioning of the implant in the anterior portion of the intervertebral space, and then through activation of the rotational pivot between the Handle and implant for final placement of the crescent-shaped cage across the anterior portion of the disc space. The AccuLIF TL Cages are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DOD) at one level or two contiguous levels from L2 to S1.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model# 800804 Lot numbers 160010 and 170003
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Howmedica Osteonics Corp.
- Reason for Recall:
- When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the disc space with a fixed engagement between the two components to permit a controlled positioning of the implant in the anterior portion of the intervertebral space, and then through activation of the rotational pivot between the Handle and implant for final placement of the crescent-shaped cage across the anterior portion of the disc space. The AccuLIF TL Cages are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DOD) at one level or two contiguous levels from L2 to S1.
Product Codes/Lot Numbers:
Model# 800804 Lot numbers 160010 and 170003
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0810-2018
Related Recalls
TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Howmedica Osteonics
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.