NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304 UDI-M535SAK3040, 5) REF SAK-306 UDI-M535SAK3060, 6) REF SAK-307 UDI- M535SAK3070, 6) REF SAK-402 UDI-M535SAK4020, 7) REF SAK-405 UDI- M535SAK4050, 8) REF SAK-406 UDI-M535SAK4060 & REF SAK-407 UDI- M535SAK4070 Hemodialysis systems and accessories
Class I - DangerousWhat Should You Do?
- Check if you have this product: SAK-301- Lot # 70879245, Exp Date 7/11/2019 SAK-302- Lot # 71079056, Exp Date 7/30/2019 SAK-303- Lot #'s 70879181 & 70879224, SAK-304- Lot #'s 70779236, 70879045, 70879105, 70879220 & 70979066, Exp Date 7/8/2019 SAK-306- Lot # 70979110, Exp Date 7/24/2019 SAK-307- Lot #'s 70879187 & 71079144, Exp Date 8/27/2019 SAK-402- Lot # 70879235, Exp Date 7/11/2019 SAK-405- Lot #'s 70779202 & 70879233, Exp Date 7/9/2019 SAK-406- Lot #'s 70979253, Exp Date 7/29/2019 SAK-407- Lot #'s 70879236 , Exp Date 7/11/2019
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- NxStage Medical, Inc.
- Reason for Recall:
- May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to prepare product.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304 UDI-M535SAK3040, 5) REF SAK-306 UDI-M535SAK3060, 6) REF SAK-307 UDI- M535SAK3070, 6) REF SAK-402 UDI-M535SAK4020, 7) REF SAK-405 UDI- M535SAK4050, 8) REF SAK-406 UDI-M535SAK4060 & REF SAK-407 UDI- M535SAK4070 Hemodialysis systems and accessories
Product Codes/Lot Numbers:
SAK-301- Lot # 70879245, Exp Date 7/11/2019 SAK-302- Lot # 71079056, Exp Date 7/30/2019 SAK-303- Lot #'s 70879181 & 70879224, SAK-304- Lot #'s 70779236, 70879045, 70879105, 70879220 & 70979066, Exp Date 7/8/2019 SAK-306- Lot # 70979110, Exp Date 7/24/2019 SAK-307- Lot #'s 70879187 & 71079144, Exp Date 8/27/2019 SAK-402- Lot # 70879235, Exp Date 7/11/2019 SAK-405- Lot #'s 70779202 & 70879233, Exp Date 7/9/2019 SAK-406- Lot #'s 70979253, Exp Date 7/29/2019 SAK-407- Lot #'s 70879236 , Exp Date 7/11/2019
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0811-2018
Related Recalls
NxStage PureFlow-B Solution. Catalog numbers: RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-406, RFP-453, RFP-454, RFP-456.
NxStage Medical
Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
The smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
There is a potential patient health risk while performing chronic hemodialysis treatments in the home setting with for patients with low blood volume (2.4 Liters or less in a patient with body weight of 30kg or less) when prescribing larger dialysate volumes.