Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
ā ļø Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer, Inc.
- Reason for Recall:
- The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
Product Codes/Lot Numbers:
Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754
Distribution:
Distributed in: MN, NJ, NC, PA, AL, MI, OR, FL, OH, CO, IL, CA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0853-2015
Related Recalls
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.