LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutting Block Size 6, Product Number 71434411, Lot Numbers: 17LM05397, 17LM05398, 17LM07849; LEGION AP Cutting Block Size 7, Product Number 71434412, Lot Numbers: 17LM01746, 17LM05390, 17LM05394, 17LM07851; LEGION AP Cutting Block Size 8, Product Number 71434413, Lot Number: 17LM08442.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Smith & Nephew, Inc.
Reason for Recall:
Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Product Codes/Lot Numbers:

LEGION AP Cutting Block Size 4, Product Number 71434409, Lot Numbers: 17LM05399, 17LM07852, 17LM07852A; LEGION AP Cutting Block Size 6, Product Number 71434411, Lot Numbers: 17LM05397, 17LM05398, 17LM07849; LEGION AP Cutting Block Size 7, Product Number 71434412, Lot Numbers: 17LM01746, 17LM05390, 17LM05394, 17LM07851; LEGION AP Cutting Block Size 8, Product Number 71434413, Lot Number: 17LM08442.

Distribution:

Distributed in: IA, NC, PA, TN, CH, BE, GB, AE, CN, DE, NL, BR, FR

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0878-2020

Related Recalls

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

Mar 18, 2025 Other Medical Devices Nationwide View Details →