BD Regional Block Needles 22G x 1.50in (50 units/cs.). Indicated for administration of medications to affect anesthesia or analgesia to a specific anatomical location (region). Catalog No. 408348

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All lots within expiration date UDI: (01)50382904083484
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Becton Dickinson & Company
Reason for Recall:
BD was advised by the European Notified Body, NSAI, that the instructions for use (IFU) needs to be updated to include the general contraindications
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

BD Regional Block Needles 22G x 1.50in (50 units/cs.). Indicated for administration of medications to affect anesthesia or analgesia to a specific anatomical location (region). Catalog No. 408348

Product Codes/Lot Numbers:

All lots within expiration date UDI: (01)50382904083484

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0924-2021

Related Recalls

An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.