IMMULITE 2000 Thyroglobulin (OUS). For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # Catalog # L2KTY2 Siemens Material Number (SMN): 10381648
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: 00630414962252 Kit Lots: 431, 432, 433, 434, 435, and 436.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe increased imprecision with low level patient samples
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
IMMULITE 2000 Thyroglobulin (OUS). For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # Catalog # L2KTY2 Siemens Material Number (SMN): 10381648
Product Codes/Lot Numbers:
UDI: 00630414962252 Kit Lots: 431, 432, 433, 434, 435, and 436.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0950-2023
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.