Curity All Purpose Sponges, Non-woven, 4 Ply, 4" x 4" (10.2 cm x 10.2 cm) Product Code: 8044 - Product Usage: Ideal for applying ointments, prepping, wiping needles, cleaning slides and more Rayon/polyester blend construction Highly absorbent.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Number: 18E190262 Expiration Date: 05/2023
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Cardinal Health 200, LLC
- Reason for Recall:
- Curity All Purpose Sponges were not sterilized and distributed
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Curity All Purpose Sponges, Non-woven, 4 Ply, 4" x 4" (10.2 cm x 10.2 cm) Product Code: 8044 - Product Usage: Ideal for applying ointments, prepping, wiping needles, cleaning slides and more Rayon/polyester blend construction Highly absorbent.
Product Codes/Lot Numbers:
Lot Number: 18E190262 Expiration Date: 05/2023
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0957-2019
Related Recalls
Webcol Alcohol Prep pad (70% isopropyl alcohol) 2 ply, 200 count boxes, Cardinal Health 200 LLC 3651 Birchwood Drive, Waukegan, IL 60085.
Cardinal Health 200
Non-Sterility: microbial contamination identified as Paenibacillus phoenicis
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Affected gowns have the incorrect expiration date on the product packaging. Product has a shelf-life of three years which is mislabeled on the product packaging as five (5) years.