Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI 15099590230623, Lot Numbers: 125533, 125589, 233656, 233919, 234055, 234339, 234388
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter, Inc.
- Reason for Recall:
- There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410
Product Codes/Lot Numbers:
UDI/DI 15099590230623, Lot Numbers: 125533, 125589, 233656, 233919, 234055, 234339, 234388
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0981-2024
Related Recalls
It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Beckman Coulter
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.