PrisMax System, Product Code 955724
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 07332414126018, All Serial Numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Baxter Healthcare Corporation
- Reason for Recall:
- This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
PrisMax System, Product Code 955724
Product Codes/Lot Numbers:
GTIN 07332414126018, All Serial Numbers
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0992-2022
Related Recalls
Pumps were released without full testing being performed, including occlusion alarm testing.
Sterile Water for Irrigation, 250 mL, Single-Dose Plastic Bottle, Rx Only, Baxter Healthcare Corporation Deerfield, IL 60015 NDC: 0338-0004-02.
Baxter Healthcare
Presence of particulate matter: plastic particles from the bottle rim were observed floating into the solution
SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2
Baxter Healthcare
Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.