Vasoshield Syringe Packs, Model VH-5001

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 00607567700468; Serial Numbers: 25147126, 25151632, 25152389, 25152964, 25163138, 25162077
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Maquet Cardiovascular, LLC
Reason for Recall:
Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Vasoshield Syringe Packs, Model VH-5001

Product Codes/Lot Numbers:

UDI-DI: 00607567700468; Serial Numbers: 25147126, 25151632, 25152389, 25152964, 25163138, 25162077

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0996-2023

Related Recalls

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.