CDI H/S Cuvette, non-sterile, Manufactured by Terumo Cardiovascular Group, Perfusion Pack or Smart Perfusion Pack. Product Usage: Heart/Lung Perfusion Pack is either a customized tubing pack built to user specifications or a stock tubing pack, designed for use during surgery requiring cardiopulmonary bypass. It is not designed for long-term use (greater than six hours). Perfusion Packs labeled as SMART¿ contain components that have an additive to improve blood compatibility. Non-SMART packs may contain some SMART tubing and/or connectors depending on the customer requirements.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog Number 6922, 1/2 inch -lots TE27, TH06 and TK03 Catalog Number 6923, 3/8 inch - lots TE13, TE18, TG15, TH20 and TK17 b. Model Number(s): 020204612, 020208807, 020210805, 020211806, 020493806, 044002300, 044003200, 044004102, 044004700, 044010200, 044016900, 044017601, 044018700, 044019901, 044027700, 044035101, 075104103, 075104404, 020135802, 088511700, 084103204, 047000500, 046001601, 065255600, 066110500, 075203800, 627265002, 627304502, 084118101, 020277801, 075101900, 627309201, 075104800, 627138002, 627241801, 627305101, 627239301, 044006801, 627347601, 075203501, 627232601, 627123006, 627280901, 088523400, 627282602, 044022001, 627252001, 044020900, 627147304, 627147403, 627153904, 046003400, 627308201, 044013600, 088511801, 084515200, 044009000, 627282001, 044022000, 066109900, 627286901, 075201201, 075202000, 084103204, 088504901, 088508000, 088513300, 088513501, 627133104, 627139306, 627165604, 627166101, 627167301, 627167501, 627167601, 627167701, 627168001, 627168101, 627179903, 627190501, 627190801, 627197802, 627198001, 627198102, 627212201, 627297401, 047002300, 044000600, 627251001, 627254501, 627283501, 627283601, 627321201, 044003901, 088504400, 627198903, 627214003, 044028900, 627231801, 627240802, 627243501, 627243601, 627243701, 627244602, 627244701, 627246002, 627252001, 627256901, 627260701, 627262303, 627266301, 627269502, 627286802, 627286901, 627290001, 627303901, 627304901, 627336301, 627336402, 627343801, 627346601, 627346701, 627349401 Catalog Number 6924, 1/4 inch - lots TE20, TF11, TG01 , TG15, TH 13 and TH27
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Sorin Group USA, Inc.
- Reason for Recall:
- The CDI System 500 monitor displays the "H/S DISCONNECT AT CUVETTE" error message when the CDI HIS Cuvette does not make a proper connection with the CDI H/S Probe. When this occurs, blood parameter values for HCT, Hgb and S02 do not display.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CDI H/S Cuvette, non-sterile, Manufactured by Terumo Cardiovascular Group, Perfusion Pack or Smart Perfusion Pack. Product Usage: Heart/Lung Perfusion Pack is either a customized tubing pack built to user specifications or a stock tubing pack, designed for use during surgery requiring cardiopulmonary bypass. It is not designed for long-term use (greater than six hours). Perfusion Packs labeled as SMART¿ contain components that have an additive to improve blood compatibility. Non-SMART packs may contain some SMART tubing and/or connectors depending on the customer requirements.
Product Codes/Lot Numbers:
Catalog Number 6922, 1/2 inch -lots TE27, TH06 and TK03 Catalog Number 6923, 3/8 inch - lots TE13, TE18, TG15, TH20 and TK17 b. Model Number(s): 020204612, 020208807, 020210805, 020211806, 020493806, 044002300, 044003200, 044004102, 044004700, 044010200, 044016900, 044017601, 044018700, 044019901, 044027700, 044035101, 075104103, 075104404, 020135802, 088511700, 084103204, 047000500, 046001601, 065255600, 066110500, 075203800, 627265002, 627304502, 084118101, 020277801, 075101900, 627309201, 075104800, 627138002, 627241801, 627305101, 627239301, 044006801, 627347601, 075203501, 627232601, 627123006, 627280901, 088523400, 627282602, 044022001, 627252001, 044020900, 627147304, 627147403, 627153904, 046003400, 627308201, 044013600, 088511801, 084515200, 044009000, 627282001, 044022000, 066109900, 627286901, 075201201, 075202000, 084103204, 088504901, 088508000, 088513300, 088513501, 627133104, 627139306, 627165604, 627166101, 627167301, 627167501, 627167601, 627167701, 627168001, 627168101, 627179903, 627190501, 627190801, 627197802, 627198001, 627198102, 627212201, 627297401, 047002300, 044000600, 627251001, 627254501, 627283501, 627283601, 627321201, 044003901, 088504400, 627198903, 627214003, 044028900, 627231801, 627240802, 627243501, 627243601, 627243701, 627244602, 627244701, 627246002, 627252001, 627256901, 627260701, 627262303, 627266301, 627269502, 627286802, 627286901, 627290001, 627303901, 627304901, 627336301, 627336402, 627343801, 627346601, 627346701, 627349401 Catalog Number 6924, 1/4 inch - lots TE20, TF11, TG01 , TG15, TH 13 and TH27
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1018-2016
Related Recalls
A specific lot of the FlexFlowTM Venous Cannula, 200-200, is being recalled because it is sharper or pointier than expected.
Identification of excess plastic on the tip of the cannula.
SORIN GROUP, SMART PERFUSION PACK, SMART 1/4" PEDS PACK, REF 627262502, STERILE EO, Rx Only, 1 EA
Sorin Group USA
Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.