Apollo DRF: Model: 9784220831 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 08090531 09090591 10040680 15051450 11020798
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Villa Radiology Systems LLC
- Reason for Recall:
- The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Apollo DRF: Model: 9784220831 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Product Codes/Lot Numbers:
08090531 09090591 10040680 15051450 11020798
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1065-2017
Related Recalls
Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Villa Radiology Systems
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Villa Radiology Systems
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
Apollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Villa Radiology Systems
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.