AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All units in the AutoMate 2500 Family are affected; Serial Numbers - 1107, 4048, 4055, 4025, 4052, 4079, 4091, 4076, 4072, 4073, 4074, 4111, 4097, 1054, 1055, 1048, 1049, 1147, ODL25125-1342, ODL25125-1354, 1251, ODL25255-4530, ODL25255-4531, 4126, 4127, 4169, 4171, 4341, 4324, ODL25255-4536, ODL25255-4535, ODL25255-4397, 4141, 4143, 4144, 4146, ODL25255-4606, 4148, ODL25255-4577, ODL25255-4605, 4302, 4304, 4352, ODL25255-4352, 4330, ODL25255-4419, ODL25255-4534, ODL25255-4416, ODL25255-4401, 4168, 4107, 4255, 4257, 4252, 4256, 4254, 4253, 4236, ODL25255-4437, ODL25255-4443, ODL25255-4444, ODL25255-4438, 4294, 4293, ODL25255-4556, ODL25255-4382, ODL25255-4378, ODL25255-4377, ODL25255-4379, ODL25255-4380, ODL25255-4381, ODL25255-4420, ODL25255-4421
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter Inc.
- Reason for Recall:
- A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
AutoMate 2550 "High Speed" Catalog Number ODL25255; AutoMate 1250 "Standard" Catalog Number ODL25125. Chemistry analyzer sample processing and sorting system.
Product Codes/Lot Numbers:
All units in the AutoMate 2500 Family are affected; Serial Numbers - 1107, 4048, 4055, 4025, 4052, 4079, 4091, 4076, 4072, 4073, 4074, 4111, 4097, 1054, 1055, 1048, 1049, 1147, ODL25125-1342, ODL25125-1354, 1251, ODL25255-4530, ODL25255-4531, 4126, 4127, 4169, 4171, 4341, 4324, ODL25255-4536, ODL25255-4535, ODL25255-4397, 4141, 4143, 4144, 4146, ODL25255-4606, 4148, ODL25255-4577, ODL25255-4605, 4302, 4304, 4352, ODL25255-4352, 4330, ODL25255-4419, ODL25255-4534, ODL25255-4416, ODL25255-4401, 4168, 4107, 4255, 4257, 4252, 4256, 4254, 4253, 4236, ODL25255-4437, ODL25255-4443, ODL25255-4444, ODL25255-4438, 4294, 4293, ODL25255-4556, ODL25255-4382, ODL25255-4378, ODL25255-4377, ODL25255-4379, ODL25255-4380, ODL25255-4381, ODL25255-4420, ODL25255-4421
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1151-2018
Related Recalls
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
MicroScan Neg MIC 3J REF C54814
Beckman Coulter
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.