VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Ventec Life Systems, Inc.
Reason for Recall:
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

VOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002

Product Codes/Lot Numbers:

REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655

Distribution:

Distributed in: US, CA, NY, MO

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1196-2025

Related Recalls

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Jan 9, 2024 Patient Monitors View Details →

Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.

May 10, 2021 Patient Monitors Nationwide View Details →