epoc Host 2 as follows: Product SMN epoc NXS Host, US 11413475 epoc NXS Host, EU 11413497 epoc NXS Host, Japan 11413498 epoc NXS Host, Canada 11413506 epoc NXS Host, ROW 11413518 epoc NXS Host, China-Korea 11413583

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Host Software Version /Sensor Configuration: 4.10.6/41.1 Product Description Siemens Material Number (SMN) (UDI-DI) epoc NXS Host, US 11413475 00630414606095; epoc NXS Host, EU 11413497 00630414605760; epoc NXS Host, Japan 11413498 00630414605814; epoc NXS Host, Canada 11413506 00630414605821; epoc NXS Host, ROW 11413518 00630414605678; epoc NXS Host, China-Korea 11413583 00630414612447 .
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Healthcare Diagnostics Inc
Reason for Recall:
pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

epoc Host 2 as follows: Product SMN epoc NXS Host, US 11413475 epoc NXS Host, EU 11413497 epoc NXS Host, Japan 11413498 epoc NXS Host, Canada 11413506 epoc NXS Host, ROW 11413518 epoc NXS Host, China-Korea 11413583

Product Codes/Lot Numbers:

Host Software Version /Sensor Configuration: 4.10.6/41.1 Product Description Siemens Material Number (SMN) (UDI-DI) epoc NXS Host, US 11413475 00630414606095; epoc NXS Host, EU 11413497 00630414605760; epoc NXS Host, Japan 11413498 00630414605814; epoc NXS Host, Canada 11413506 00630414605821; epoc NXS Host, ROW 11413518 00630414605678; epoc NXS Host, China-Korea 11413583 00630414612447 .

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1215-2023

Related Recalls

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Other Medical Devices Nationwide View Details →

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Infusion Pumps Nationwide View Details →

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Oct 6, 2025 Other Medical Devices Nationwide View Details →