Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Olympus Corporation of the Americas
- Reason for Recall:
- Olympus has received complaints regarding the breakage of the endoscope's insertion tube bending section during surgical procedures. Some of these complaints are associated with tissue trauma, including perforation, and insertion tubes which were stuck inside the patient and had to be surgically removed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Uretero-reno videoscope URF-V2 and URF-V2R endoscope and accessories The URF-V2/V2R endoscopes are intended for use in endoscopic diagnosis and treatment within the ureter and kidney.
Product Codes/Lot Numbers:
All serial numbers
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1252-2017
Related Recalls
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number: TB2-0545FC.
Olympus Corporation of the Americas
Potential for detachment of a distal tip component of the device during use.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.
Olympus Corporation of the Americas
XXX
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 20cm. Model Number: TB2-0520FC.
Olympus Corporation of the Americas
Potential for detachment of a distal tip component of the device during use.