MicroScan LabPro Information Manager System, Software Versions 1.0 to 4.41 Catalog No. 10638819, 10638820, 10638823, 10638824, 10638825,10638826, 10714149,10714159, 10805072, 10805073, 10975000, 10975001, 6000-0008, 6000-0026. Japan: 10638819 LabPro v4.11 Software Update Kit 10638820 LabPro v4.11 System Software 10805072 LabPro MBT Non-Connect Systems v4.30 10805073 LabPro MBT Connect Systems v4.30 Canada: 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit Europe: 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software USA: 10714149, LabPro v4.11 Software Update Kit 10714150 LabPro v4.11 System Software 10975000 ASM LabPro-MBT v4.40 10975001 ASM LabPro-MBT v4.40 for Connect 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit ROW: 6000-0008 LabPro v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software Product Usage: LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel. LabPro AlertEx is a functional subset of the LabPro Data Management System that analyzes MicroScan ID and AST data, or other predefined parameters, against a series of established rules/alerts and notifies the user of unusual and/or critical conditions, which may warrant further analysis or actions.
Class I - DangerousWhat Should You Do?
- Check if you have this product: not applicable
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter Inc.
- Reason for Recall:
- Beckman Coulter is recalling the MicroScan LabPro Information Manager System because the software incorrectly allows the operator to manually edit the carbohydrate substrates when manually reading dried overnight gram negative panels with an ID Hold status.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MicroScan LabPro Information Manager System, Software Versions 1.0 to 4.41 Catalog No. 10638819, 10638820, 10638823, 10638824, 10638825,10638826, 10714149,10714159, 10805072, 10805073, 10975000, 10975001, 6000-0008, 6000-0026. Japan: 10638819 LabPro v4.11 Software Update Kit 10638820 LabPro v4.11 System Software 10805072 LabPro MBT Non-Connect Systems v4.30 10805073 LabPro MBT Connect Systems v4.30 Canada: 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit Europe: 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software USA: 10714149, LabPro v4.11 Software Update Kit 10714150 LabPro v4.11 System Software 10975000 ASM LabPro-MBT v4.40 10975001 ASM LabPro-MBT v4.40 for Connect 6000-0008 LabPro v4.41 Panel Update-06 Kit 6000-0026 LabPro Connect v4.41 Panel Update-06 Kit ROW: 6000-0008 LabPro v4.41 Panel Update-06 Kit 10638823 LabPro v4.11 Software Update Kit 10638825 LabPro v4.11 System Software 10638824 LabPro v4.11 Software Update Kit 10638826 LabPro v4.11 System Software Product Usage: LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instruments or manually entered microbiology test results, for use by trained laboratory personnel. LabPro AlertEx is a functional subset of the LabPro Data Management System that analyzes MicroScan ID and AST data, or other predefined parameters, against a series of established rules/alerts and notifies the user of unusual and/or critical conditions, which may warrant further analysis or actions.
Product Codes/Lot Numbers:
not applicable
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1269-2016
Related Recalls
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
MicroScan Neg MIC 3J REF C54814
Beckman Coulter
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.