Colonovideoscope Model CF-H180AL

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 04953170339615 Serial number: 2501131
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Olympus Corporation of the Americas
Reason for Recall:
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Colonovideoscope Model CF-H180AL

Product Codes/Lot Numbers:

UDI-DI: 04953170339615 Serial number: 2501131

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1274-2022