Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449; Serial Numbers: CA237026J3 CH231276C6
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Datascope Corp.
Reason for Recall:
There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Product Codes/Lot Numbers:

Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449; Serial Numbers: CA237026J3 CH231276C6

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1277-2023