Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number 6 ENT
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Seiler Instr. & Mfg. Co. Inc.
- Reason for Recall:
- The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Product Codes/Lot Numbers:
Model Number 6 ENT
Distribution:
Distributed in: US, CA, CO, FL, IL, MN, NV, OK, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1294-2018