Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    This correction is applicable when TEG 5000 PlateletMapping (Item # 07-014, 07-015, 07-016) is performed on a TEG 5000 analyzer (Item # 07-022, 07-033) that is connected to TEG Manager Software which includes viewing of TEG 5000 results. Code Info: UDI-DI: (01)30812747018184 Software Version: TEGM 1.1 and later, TAS v4.2.3
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Haemonetics Corporation
Reason for Recall:
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Product Name: TEG5000 Analyzer (07-022, 07-033) with TEG Analytical Software (07-030, 07-031) and TEG Platelet Mapping Assay - PlateletMapping ADP & AA (07-014), PlateletMapping ADP (07-015), PlateletMapping AA (07-016). Issue occurs when Platelet Mapping Assay results from a TEG 5000 analyzer are viewed using TEG Manager Software (versions 1.1-4.3.1)

Product Codes/Lot Numbers:

This correction is applicable when TEG 5000 PlateletMapping (Item # 07-014, 07-015, 07-016) is performed on a TEG 5000 analyzer (Item # 07-022, 07-033) that is connected to TEG Manager Software which includes viewing of TEG 5000 results. Code Info: UDI-DI: (01)30812747018184 Software Version: TEGM 1.1 and later, TAS v4.2.3

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1298-2022

Related Recalls

Due to a shift in the citrated blood reference range for a normal population, the ranges stated in the IFU are changing. If the clinician relies solely on affected results in comparison to the normal donor reference range, it could lead to misdiagnosis and incorrect treatment or failure to treat.

Apr 26, 2022 Other Medical Devices Nationwide View Details →

Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.

Feb 19, 2020 Other Medical Devices Nationwide View Details →

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.