Dynavox T15 Speech Generating Device ( including Batteries)
Class I - DangerousWhat Should You Do?
- Check if you have this product: TL1405 TL1407 Part Numbers: T15: 750423 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Dynavox Systems Llc
- Reason for Recall:
- The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Dynavox T15 Speech Generating Device ( including Batteries)
Product Codes/Lot Numbers:
TL1405 TL1407 Part Numbers: T15: 750423 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1305-2018
Related Recalls
Dynavox T10 Speech Generating Device ( including Batteries)
Dynavox Systems
The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.