CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Class I - DangerousWhat Should You Do?
- Check if you have this product: (domestic only) SN374 SN17081704 SN17121782 SN18011800 SN18081862 SN18061846 SN18111907
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems Gmbh, DMC
- Reason for Recall:
- When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Product Codes/Lot Numbers:
(domestic only) SN374 SN17081704 SN17121782 SN18011800 SN18081862 SN18061846 SN18111907
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1329-2020
Related Recalls
CombiDiagnost R90 Software Version R1.0 and R1.1
Philips Medical Systems Gmbh, DMC
Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Philips Medical Systems Gmbh, DMC
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
Stationary Fluoroscopic X-ray System CombiDiagnost R90, Model 709030 Stationary Fluoroscopic X-ray System
Philips Medical Systems Gmbh, DMC
kV/mA lockin not functioning as specified which may result in the patient receiving an increased radiation dose.