Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: WAA090630, WAA090724, WAA091212, WAA091456, WAA091468, WAA092596. UDI Code 07340153700031
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Maquet Cardiovascular Us Sales, Llc
- Reason for Recall:
- Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-502 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.
Product Codes/Lot Numbers:
Serial Numbers: WAA090630, WAA090724, WAA091212, WAA091456, WAA091468, WAA092596. UDI Code 07340153700031
Distribution:
Distributed in: CA, CO, IA, MN, NY, SC, VA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1390-2020
Related Recalls
HLS Set Advanced Model Name: BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code/Part Number: 70105.2797 - Product Usage: These devices are used with the CARDIOHELP system in order to provide extracorporeal circulation for cardiac support and/or pulmonary support.
Maquet Cardiovascular Us Sales
The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product's sterility barrier.
HLS Set Advanced, Model: BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Product Code/Part Number: 70105.2794 - Product Usage: These devices are used with the CARDIOHELP system in order to provide extracorporeal circulation for cardiac support and/or pulmonary support.
Maquet Cardiovascular Us Sales
The HLS Set Advanced (disposable for Cardiohelp) is being recalled due to breaches in the product's sterility barrier.
The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.