Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 83010 85001 85002 83005-30 83005-35 83005-40 83005-45 83005-50 83006-30 83006-35 83006-40 83006-45 83006-50 83006-55 83006-60 83007-30 83007-35 83007-40 83007-45 83007-50 83007-55 83008-30 83008-35 83008-40 83008-45 83008-50 83008-55 83008-80 82001-1 Catalog Numbers: LA-HSS0010 LA-HDOO01 LA-CTOO02 LA-PS0530 LA-PS0535 LA-PS0540 LA-PS0545 LA-PS0550 LA-PS0630 LA-PS0635 LA-PS0640 LA-PS0645 LA-PS0650 LA-PS0655 LA-PS0660 LA-PS0730 LA-PS0735 LA-PS0740 LA-PS0745 LA-PS0750 LA-PS0755 LA-PS0830 LA-PS0835 LA-PS0840 LA-PS0845 LA-PS0850 LA-PS0855 LA-PS0880 LA-SOO01 Lots Affected: All
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Alphatec Spine, Inc.
- Reason for Recall:
- Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head and rod may fail to provide a rigid construct under loading. To date there have been no reports of patient injury, adve
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments.
Product Codes/Lot Numbers:
83010 85001 85002 83005-30 83005-35 83005-40 83005-45 83005-50 83006-30 83006-35 83006-40 83006-45 83006-50 83006-55 83006-60 83007-30 83007-35 83007-40 83007-45 83007-50 83007-55 83008-30 83008-35 83008-40 83008-45 83008-50 83008-55 83008-80 82001-1 Catalog Numbers: LA-HSS0010 LA-HDOO01 LA-CTOO02 LA-PS0530 LA-PS0535 LA-PS0540 LA-PS0545 LA-PS0550 LA-PS0630 LA-PS0635 LA-PS0640 LA-PS0645 LA-PS0650 LA-PS0655 LA-PS0660 LA-PS0730 LA-PS0735 LA-PS0740 LA-PS0745 LA-PS0750 LA-PS0755 LA-PS0830 LA-PS0835 LA-PS0840 LA-PS0845 LA-PS0850 LA-PS0855 LA-PS0880 LA-SOO01 Lots Affected: All
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1421-2013
Related Recalls
Complaints have been received regarding post-operative implant collapse.
Due to reports of intraoperative graft bolt implantation breakages.
Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.