Apollo EZ; Model: 9784152036 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number: 13101261
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Villa Sistemi Medicali S.P.A.
Reason for Recall:
Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Apollo EZ; Model: 9784152036 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures

Product Codes/Lot Numbers:

Serial Number: 13101261

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1430-2017

Related Recalls

An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.

Nov 4, 2019 Other Medical Devices Nationwide View Details →

An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.

Nov 4, 2019 Other Medical Devices Nationwide View Details →

An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.

Nov 4, 2019 Other Medical Devices Nationwide View Details →