Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: SIGSDS30CTVT; UDI/DI: 10884521741881; Lot number: N3L1909UY;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Covidien, LP
- Reason for Recall:
- Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Product Codes/Lot Numbers:
Model Number: SIGSDS30CTVT; UDI/DI: 10884521741881; Lot number: N3L1909UY;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1457-2024
Related Recalls
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting