Terumo Advanced Perfusion System 1-Level Sensor II Gel Pads Product Usage: Level sensor pads are used to attach the level sensors to a hard shell reservoir. includes coupling gel.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Level Sensor II Gel Pads, Catalog No. 217390, UDI No: 00886799000519, Lot Numers Ranging from 782300 through 817488, Manufactured from 19-Nov-2015 to 25-Nov-2016; Distributed from 23-Nov-2015 through 15-Dec-2016.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Terumo Cardiovascular Systems Corporation
- Reason for Recall:
- Terumo CVS initiated a voluntary recall for the Level Sensor II Pads and Level Sensor Gel Pads due to non-compliant labeling because the product expiration date is displayed in a format that may not be recognizable to all users.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Terumo Advanced Perfusion System 1-Level Sensor II Gel Pads Product Usage: Level sensor pads are used to attach the level sensors to a hard shell reservoir. includes coupling gel.
Product Codes/Lot Numbers:
Level Sensor II Gel Pads, Catalog No. 217390, UDI No: 00886799000519, Lot Numers Ranging from 782300 through 817488, Manufactured from 19-Nov-2015 to 25-Nov-2016; Distributed from 23-Nov-2015 through 15-Dec-2016.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1459-2017
Related Recalls
CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit
Terumo Cardiovascular Systems
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Terumo Cardiovascular Systems
During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention