Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP¿ platform.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318). UDI Code: (01) 1 0627966 00508 4 (10) 070318 (17)300919
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Response Biomedical Corp.
- Reason for Recall:
- The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP¿ platform.
Product Codes/Lot Numbers:
Randox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318). UDI Code: (01) 1 0627966 00508 4 (10) 070318 (17)300919
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1461-2020