Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial no: ASKF-0224 ASKF-0232 ASKH-0004 ASKH-0007 ASKH-0038 ASKH-0088 ASKH-0132 ASKH-0133 ASKJ-0007 ASKJ-0178 ASKJ-0179 ASKJ-0184 ASKJ-0205 ASKJ-0220 ASKJ-0224 ASKK-0002 ASKK-0006 ASKK-0010 ASKK-0011 ASKK-0014 ASKK-0019 ASKK-0022 ASKK-0023 ASKK-0032 ASKK-0036 ASKK-0038 ASKK-0044
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Draeger Medical, Inc.
- Reason for Recall:
- The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Fabius GS Premium, anesthesia machine, catalog no. 8607000 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).
Product Codes/Lot Numbers:
Serial no: ASKF-0224 ASKF-0232 ASKH-0004 ASKH-0007 ASKH-0038 ASKH-0088 ASKH-0132 ASKH-0133 ASKJ-0007 ASKJ-0178 ASKJ-0179 ASKJ-0184 ASKJ-0205 ASKJ-0220 ASKJ-0224 ASKK-0002 ASKK-0006 ASKK-0010 ASKK-0011 ASKK-0014 ASKK-0019 ASKK-0022 ASKK-0023 ASKK-0032 ASKK-0036 ASKK-0038 ASKK-0044
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1482-2018
Related Recalls
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Draeger Medical
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.