3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model 5XL, Serial Numbers: 722612 - 722683, 820352 - 820360 Model 8XL, Serial Numbers: 351303 - 351399, 450144 - 450147
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
3M Company - Health Care Business
Reason for Recall:
3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.

Product Codes/Lot Numbers:

Model 5XL, Serial Numbers: 722612 - 722683, 820352 - 820360 Model 8XL, Serial Numbers: 351303 - 351399, 450144 - 450147

Distribution:

Distributed in: AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1563-2016