Alinity m System, Part Number: 08N53-002
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: 08N53-002; UD/DI: 00884999048034. Not Serial Number specific
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Molecular, Inc.
- Reason for Recall:
- The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Alinity m System, Part Number: 08N53-002
Product Codes/Lot Numbers:
Part Number: 08N53-002; UD/DI: 00884999048034. Not Serial Number specific
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1602-2024
Related Recalls
There were reports of an increase in reactive negative controls and false positive results.
There were reports of an increase in reactive negative controls and false positive results.
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.