MultiDiagnostic-Eleva with Flat Detector 708035 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Pending
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Medical Systems Nederlands
Reason for Recall:
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

MultiDiagnostic-Eleva with Flat Detector 708035 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Product Codes/Lot Numbers:

Pending

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1609-2018

Related Recalls

Certain low-voltage DC power supplies (DCPS) used in these products may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering.

Jun 23, 2016 Other Medical Devices View Details →
Class I - Dangerous

Philips has discovered through customer complaints and internal testing an intermittent electronic product defect in accordance with 21 CFR 1003.2(b)(1). When performing Cine runs, the Azurion's automatic exposure control software may set technique factors that result in radiation emissions too low to obtain useful diagnostic images.

Nov 30, 2018 Other Medical Devices Nationwide View Details →

Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling.

Mar 25, 2019 Diagnostic Equipment View Details →