Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of cancer
Class I - DangerousWhat Should You Do?
- Check if you have this product: material numbers and serial numbers: 1940035 2962 8139789 5833 8139789 5899 9401654 2113 1940035 2628 4504200 3066 4504200 3361 1940035 3873 1940035 2855
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- Siemens Radiation Oncology became aware that customers may be using the Siemens Linear Accelerator in combination with stereotactic accessories which have not been validated as being compatible with Siemens LINAC models.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Siemens Linear Accelerator (LINAC) models. Product Usage: deliver X-ray radiation for therapeutic treatment of cancer
Product Codes/Lot Numbers:
material numbers and serial numbers: 1940035 2962 8139789 5833 8139789 5899 9401654 2113 1940035 2628 4504200 3066 4504200 3361 1940035 3873 1940035 2855
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1652-2014
Related Recalls
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS Pheno VE30A and VE40A, Model 10849000
Siemens Medical Solutions USA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.