Quadrox-iD Adult Diffusion Membrane Oxygenator Product Usage: The Quadrox-iD Adult Diffusion Membrane Oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The oxygenator with diffusion membrane cannot be used for the administration of volatile anesthetic gases because the gas molecules are unable to pass through the diffusion membrane. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Part numbers BO-HMOD 70000, 70105.3815, BEQ-HMOD 70000, 70105,3824
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Maquet Cardiovascular Us Sales, Llc
Reason for Recall:
A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and Diffusion Membrane HLS Set Advanced and HIT Set Advanced Models.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Quadrox-iD Adult Diffusion Membrane Oxygenator Product Usage: The Quadrox-iD Adult Diffusion Membrane Oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The oxygenator with diffusion membrane cannot be used for the administration of volatile anesthetic gases because the gas molecules are unable to pass through the diffusion membrane. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Product Codes/Lot Numbers:

Part numbers BO-HMOD 70000, 70105.3815, BEQ-HMOD 70000, 70105,3824

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1689-2015

Related Recalls

The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.

Sep 30, 2020 Diagnostic Equipment Nationwide View Details →