Magic3 Go Intermittent Urinary Catheters, pediatric Catalog Numbers: a) REF 52810G

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a) REF 52810G, UDI/DI 00801741136870, Lot Numbers: JUGW1445
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
C.R. Bard Inc
Reason for Recall:
Some catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Magic3 Go Intermittent Urinary Catheters, pediatric Catalog Numbers: a) REF 52810G

Product Codes/Lot Numbers:

a) REF 52810G, UDI/DI 00801741136870, Lot Numbers: JUGW1445

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1700-2023