EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 5040207003 5040207004 5040308003 5042100008 5042100016 5042100017 5042100018 5042100019 5042100020 5042100021 5042100022 5042100023 5042100024 5042100025 5042100033 5042100034 5042100035 5042100036 5042100037 5042100038 5042100039 5042100040 5042100060 5042100061 5042100063 5042100064 5042100065 5042100066 5042100067 5042110002 5042110003 5042110004 5042110005 5042110006 5042120002 5042120006 5042130009 5042130011 5042140029 5042140030 5042140031 5042140032 5042140038 5042140039
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Advanced Sterilization Products
- Reason for Recall:
- Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
EVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.
Product Codes/Lot Numbers:
5040207003 5040207004 5040308003 5042100008 5042100016 5042100017 5042100018 5042100019 5042100020 5042100021 5042100022 5042100023 5042100024 5042100025 5042100033 5042100034 5042100035 5042100036 5042100037 5042100038 5042100039 5042100040 5042100060 5042100061 5042100063 5042100064 5042100065 5042100066 5042100067 5042110002 5042110003 5042110004 5042110005 5042110006 5042120002 5042120006 5042130009 5042130011 5042140029 5042140030 5042140031 5042140032 5042140038 5042140039
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1715-2015
Related Recalls
STERRAD NX Sterilization System, Product Code: 10033, 10033-002
Advanced Sterilization Products
Advanced Sterilization Products (ASP) has identified that in a rare sequence of events following an external power outage, the system door interlock can be become disengaged during system reboot for a 20-second duration.
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.
CIDEX OPA Solution, P/N 20390 for use as a high level disinfectant for reprocessing heat sensitive medical devices.
Advanced Sterilization Products
Advanced Sterilization Products (ASP) is reminding all CIDEX OPA Solution customers to follow the warnings and rinsing procedure in their current labelling around the criticality of proper rinsing of medical device following disinfection with CIDEX OPA according to the CIDEX OPA Solution IFU (Instruction For Use).