HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902, 611001, 611002, 611004, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. First two digits (61) refer to model number 610 for the HLD system Second two digits - two digit year of manufacture Last two digits sequentially numbered as they receive orders.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Cenorin, LLC
- Reason for Recall:
- Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.
Product Codes/Lot Numbers:
610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902, 611001, 611002, 611004, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. First two digits (61) refer to model number 610 for the HLD system Second two digits - two digit year of manufacture Last two digits sequentially numbered as they receive orders.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1729-2014
Related Recalls
There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.
Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.