GETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 720001F0 (5) Part 710001B2 (6) Part 710001B0 (7) Part 710001F0 Product Usage: The Meera Mobile Operating Table is designed to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    (1) Part 720001B2 S/N 1 to S/N 416 (2) Part 720001B0 S/N 1 to S/N 1416 (3) Part 720001F2 S/N 1 to S/N 91 (4) Part 720001F0 S/N 1 to S/N 220 (5) Part 710001B2 S/N 1 to S/N 8 (6) Part 710001B0 S//N 1 to S/N 11 (7) Part 710001F0 S/N 1 Expanded lots: 720001B2 SN 417 to SN 503 720001B0 SN 1417 to SN 1689 720001F2 SN 92 to SN 119 720001F0 SN 221 to SN 226 and SN 701 to 711 710001B2 SN 9 to SN 14 710001B0 SN 12 to SN 88
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Maquet Cardiovascular Us Sales, Llc
Reason for Recall:
User may suffer an electric shock when coming in contact with the exposed electronic components
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

GETINGE MEERA MOBILE OPERATING TABLE: (1) Part 720001B2 (2) Part 720001B0 (3) Part 720001F2 (4) Part 720001F0 (5) Part 710001B2 (6) Part 710001B0 (7) Part 710001F0 Product Usage: The Meera Mobile Operating Table is designed to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment.

Product Codes/Lot Numbers:

(1) Part 720001B2 S/N 1 to S/N 416 (2) Part 720001B0 S/N 1 to S/N 1416 (3) Part 720001F2 S/N 1 to S/N 91 (4) Part 720001F0 S/N 1 to S/N 220 (5) Part 710001B2 S/N 1 to S/N 8 (6) Part 710001B0 S//N 1 to S/N 11 (7) Part 710001F0 S/N 1 Expanded lots: 720001B2 SN 417 to SN 503 720001B0 SN 1417 to SN 1689 720001F2 SN 92 to SN 119 720001F0 SN 221 to SN 226 and SN 701 to 711 710001B2 SN 9 to SN 14 710001B0 SN 12 to SN 88

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1734-2019

Related Recalls

The Printed Circuit Board Assembly (PCBA) are not in compliance with the IEC standard (60601 1, third edition). The BMU 40 units affected by this recall were designed with a short creepage distance of 1.8 mm between the two distinct and proximal areas on the printed circuit board (PCB) where as the standard requires a creepage distance of 4 mm.

Sep 30, 2020 Diagnostic Equipment Nationwide View Details →